Coherent FieldMaxII-TOP Manual de usuario Pagina 58

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 118
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 57
FieldMaxII-TOP User Manual
44
Measurement
Display and
Range
Selection
The display update rate for numeric measurement is
three times per second.
Measurement range is selected in decade steps. Range
selection (shown in Table 3 on page 45, and Table 4 on
page 46) is dependent on the sensor type and character-
istics, as well as user-determined measurement settings.
FieldMaxII-TOP uses the “3’s” Rule—a display format-
ting rule in which the display value is not allowed to
exceed 3, 30, or 300, depending on where the decimal
point falls with the decimal point located in a fixed posi-
tion, as determined by the current range. Typically if a
reading exceeds the “3’s” limit, the instrument is over-
ranged or, in the case of Auto Ranging, the instrument
will automatically range up.
Vista de pagina 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 117 118

Comentarios a estos manuales

Sin comentarios