Operation
49
In the example shown in Figure 15, the internal trigger
threshold has been set to 8% (shown as a dashed line).
Pulse A will definitely not generate a reliable trigger.
Pulse B may generate a trigger, but not reliably. Pulses C
and D will definitely generate reliable triggers.
A full-scale signal on one range may not trigger on the
next higher range unless the trigger level is set to less
than 10%. For example, a near full-scale signal of 280 mJ
on the 300 mJ range is less than 10% of full scale on the
next higher range (3 J), and would therefore not trigger
the reading. To obtain reliable triggering in this instance,
adjust the trigger level to less than 8%.
Figure 15. Internal Trigger Threshold