Operation
29
Stat
Pressing the Stat button will:
• Enter Statistics mode—if Statistics mode is not
active and the button is pressed for less than two
seconds.
• Exit Statistics mode.
• Enter Edit mode and select the statistics parameters
to be edited (Batch Size and Restart mode) after
edit cycle initiation (if Setup/Local is pressed
beforehand). Batch size is 1 to 99,999 seconds.
• View the statistics parameters if the button is
pressed for two seconds or more.
Wave
You can configure FieldMaxII-TO to automatically
account for any difference between the laser wave-
length and the calibration wavelength. In the case
of optical sensors, this compensation is necessary
because the sensor contains calibration data from a
number of different wavelengths. Thermopile and
optical sensors include wavelength compensation infor-
mation that is used in this mode.
After pressing Setup/Local, the Wave button is used to
enter Edit mode and set the wavelength. If an edit cycle
has not been initiated, pressing the Wave button will